Products
Coming 2026
Products
Grow With CellGenic

The Anicell is designed to culture eukaryotic cells such as CHO, HEK, Hela etc. in suspension but can also be used as a static incubator if required.
Industry Standard Dual Beam Infra Red Sensor is used to maintain control of CO2 density.
All six sides are directly heated and combined with P.I.D control to ensure that temperatures are reached quickly and uniformity is maintained. Further the system combines forced air and natural convection to maintain the best temperature uniformity at al times.
A deep and wide humidity tray allows a high and natural humidity to be generated.
The internal chamber is separated into three compartments each with its own shelve, shaker and glass door. This design prevents excessive loss of heat and CO2 when removing flasks.
An LED display of actual humidity in the chambers informs the user of the appropriate time to supplement the water in the humidity tray.
Each shaker has an auto start function such that if the glass inner door to each compartment is opened the shaker in that compartment stops and starts again when the door is closed. Auto start is also enabled if there is a power cut.
Platform of each shakers can be pulled out to load flasks efficiently on each shakers.
The orbital shakers can be used with sticky mats or dedicated flask holder trays or universal flask/tube holder plate.
Each orbital shaker can be individually controlled with an external control panel.
A UV lamp situated next to the circulation fan works to sterilize the air in the incubator even during cell culturing. The UV lamp can be turned on or off with a switch on the front panel.
The inner chamber and all orbital shakers are constructed with stainless steel SUS304 which is designed for use in GMP facilities and is resistant to rust formation in high humidity conditions.
A HEPA filter, located in the post circulation fan, traps microbes and helps to maintain a sterile environment.
| INCUBATOR | NB-206CXL | NB-206CXXL |
|---|---|---|
| Temp. range | Ambient +5℃ to 60℃ | Ambient +5℃ to 60℃ |
| Temp. accuracy | ±1℃ at 37℃ | ±1℃ at 37℃ |
| Humidity | ≥70% at 37℃ | ≥70% at 37℃ |
| CO2 range | 0% to 20% | 0% to 20% |
| CO2 accuracy | ±0.3% at 5% at 37℃ | ±0.3% at 5% at 37℃ |
| CO2 sensor | IR CO2 sensor | IR CO2 sensor |
| CO2 inlet pressure | 0.7 ~ 1 bar | 0.7 ~ 1 bar |
| Outer door | Silicon packing magnet door | Silicon packing magnet door |
| Inner door | Each inner door of 3 Shelves | Each inner door of 3 Shelves |
| Display | LED Display | LED Display |
| Jacket type | Air jacket type (6 sides heating) | Air jacket type (6 sides heating) |
| Filter | Anti-Bacterial Hepa filter | Anti-Bacterial Hepa filter |
| Sterilization | U.V 4Wx1ea | U.V 4Wx1ea |
| Chamber volume | 650 liter | 850 liter |
| Shelves | 3ea | 3ea |
| Chamber dimensions | 700(W)x650(D)x1430(H)mm | 700(W)x800(D)x1530(H) mm |
| Each compartment Dimension | Compartment 1(Bottom): 700(W)x650(D)x430(H) mm
Compartment 2, 3 (Middle, Top): 700(W)x650(D)x380(H) mm |
700(W)x800(D)x410(H) mm |
| Overall dimensions | 820(W)x780(D)x1740(H) mm | 820(W)x920(D)x1840(H) mm |
| Weight | 323kg | 393kg |
| Power | 110/220V, 50/60Hz | 110/220V, 50/60Hz |
| SHAKER | ||
| Shaking motion | Orbital | Orbital |
| Speed range | 30 to 200 rpm | 30 to 250 rpm |
| Speed accuracy | ±1rpm | ±1rpm |
| Speed increment | 1rpm | 1rpm |
| Time range | Continuous or up to 47h 59mins | Continuous or up to 47h 59mins |
| Time accuracy | ±1% | ±1% |
| Time increment | 1 minute | 1 minute |
| Motor | Plate type BL/DC Motor | Plate type BL/DC Motor |
| Drive system | Beltless direct drive | Beltless direct drive |
| Orbit diameter | 25mm | 25mm |
| Platform size | 520(W)x520(D) mm | 520(W)x520(D) mm |
| Dimension | 465(W)x540(D)x125(H) mm | 465(W)x540(D)x125(H) mm |
| Flask Capacity | NB-206CXL | NB-206CXXL |
|---|---|---|
| 125mL Flask | Max. 158ea | Max. 158ea |
| 250mL Flask | Max. 90ea | Max. 90ea |
| 500mL Flask | Max. 60ea | Max. 60ea |
| 1000mL Flask | Max. 48ea | Max. 48ea |
| 2000mL Flask | Max. 9ea | Max. 27ea |
Optimum performance for maximum allowance can be obtained at 150rpm.
| 206CXL-H10 | 100ml Flask Holder x 36ea with Platform |
| 206CXL-H25 | 250ml Flask Holder x 23ea with Platform |
| 206CXL-H50 | 500ml Flask Holder x 16ea with Platform |
| 206CXL-H100 | 1000ml Flask Holder x 9ea with Platform |
| 206CXL-H200 | 2000ml Flask Holder x 5ea with Platform |
| 206CXL-SR | Spring Rack |
| 206CXL-MM | Microplate rack with platform(6ea) |
| 206CXL-PL | Hard-Anodizing Alluminum Plate with universal hole |
| 206CXL-SLPL | SLIDE OUT PLATFORM FOR SHAKER |
Lot-traced · COA on file · Sterility & mycoplasma tested
PEPTIDE DOSE: Select dose
DRAW SYRINGE TO: Select values
YOUR VIAL CONTAINS: Select values
CONCENTRATION: Select values
The aniCell CO₂ Shaking Incubator is a high-capacity cell culture system designed to provide precise environmental control and reliable shaking performance for suspension cell culture applications. Widely used in biotechnology, proteomics, genomics, crystallography, cell biology, and pharmaceutical research, the aniCell delivers optimal conditions for advanced laboratory workflows.
The incubator features a spacious chamber divided into three independent compartments, each equipped with a removable orbital shaker capable of accommodating Erlenmeyer flasks, cylindrical flasks, and deep-well blocks. This flexible configuration allows researchers to conduct multiple culture processes simultaneously while maintaining excellent environmental consistency.
Apply once, get reviewed by our team, and access provider-only pricing across peptides, exosomes, and cellular products.
PRODUCT USAGE: This PRODUCT IS INTENDED AS A RESEARCH CHEMICAL ONLY. This designation allows the use of research chemicals strictly for in vitro testing and laboratory experimentation only. All product information available on this website is for educational purposes only. This product is intended strictly for laboratory research and qualifies under appropriate regulatory frameworks for research-only use. It is not intended to diagnose, treat, cure, or prevent any disease or condition. This product should only be handled by licensed, qualified professionals. It is not intended for use by the general public. Bodily introduction of any kind into humans or animals is strictly forbidden by law.